The AC1921T Multi-Channel AC/DC Hipot & Insulation Tester integrates AC/DC withstand voltage and insulation resistance test functions. It supports 8-channel (optional) synchronous parallel testing and cross-channel testing. Hipot test capacity is 5 kV / 100 mA; insulation resistance testing covers 100 V –6000 V / 50 GΩ. Safety test accuracy achieves 1% class.
| Model | Test Capacity | Voltage Accuracy | Resolution | Interface | Size | Inquiry | |
|---|---|---|---|---|---|---|---|
| AC1921T | 5KV/100mA | ±(1 % rdg. + 5 V) | 1V | PLC Digital I/O / RS-232 / WiFi | 3U | ||
| No models match the selected filters | |||||||
*Specifications for the product series are subject to change without notice.
• Three test capabilities: insulation resistance, AC withstand voltage and DC withstand voltage, with 1% accuracy for safety test items.
• Portable design: compact dimensions of 132 × 284 × 470 mm and a light weight of 13 kg.
• Enhanced Compliance Performance
• Withstand voltage capacity: 200 VA (5 kV / 40 mA), with DC withstand voltage test function.
• Insulation resistance test voltage range: 100 V–6000 V, measurement range up to 100 GΩ.
• Accurate measurement of ultra-low breakdown current at the μA level.
• Information Management
• Equipped with Wi-Fi, RS-232 and other communication interfaces. A single PC can connect to multiple instruments for test data upload, storage and analysis.
• Local data caching with transmission resumption after network disconnection.
• When connected to the network, it supports remote firmware updates and remote diagnostics.
• Easy Operation
• 7-inch color touchscreen with an index-style interface for convenient operation and intuitive display.
• Built-in help information facilitates operator training and onboarding;
• Two modes: Administrator and Operator. Administrators can preset parameters; operators can only use them and cannot make changes.
• Error-proof connectors for instrument accessories ensure simple installation and setup.
Product overview
The AC1921T Multi-Channel AC/DC Hipot & Insulation Tester integrates AC/DC withstand voltage and insulation resistance test functions. It supports 8-channel (optional) synchronous parallel testing and cross-channel testing. When used with appropriate test fixtures, it enables synchronous batch testing and significantly improves production efficiency. Each channel displays test results independently, with automatic judgment and audible-visual alarms for Pass, Fail and Missed Test statuses. Hipot test capacity is 5 kV / 100 mA; insulation resistance testing covers 100 V –6000 V / 50 GΩ. Safety test accuracy achieves 1% class.The instrument features a 7-inch full-color touchscreen for intuitive operation and clear display.
With overall dimensions of 132 × 284 × 470 mm, the unit is compact and portable. Its modular functional circuit design includes self-diagnostic capability, improving after-sales service efficiency. Test data is stored locally on the instrument. It provides Wi-Fi, RS-232, and other communication interfaces. It supports multi-unit networking and enables remote firmware upgrades over the network.
Model | AC1921T | ||
Product Name | AC/DC Hipot and Insulation Tester | ||
AC Withstand Voltage Test | |||
Range | Resolution | Accuracy | |
Output Voltage | 300~5000V AC | 1V | ±(1% rdg. + 5 V) |
Output Frequency | 50/60Hz | ||
Breakdown Current | 0~100.00mA | 0.001mA/0.01mA | ±(1% + 0.002mA) |
Instantaneous 200mA peak | |||
Ramp-up Time | 0~999.9s | 0.1s | ±(1% + 0.1s) |
Ramp-down Time | 0~999.9s | 0.1s | ±(1% + 0.1s) |
Test Time | 0.5~999.9s | 0.1s | ±(1% + 0.1s) |
Current Compensation | 0~99.999mA | 0.001mA/0.01mA | |
DC Withstand Voltage Test | |||
Range | Resolution | Accuracy | |
Output Voltage | 300~6000VDC | 1V | ±(1% rdg. + 5 V) |
Breakdown Current | 0~9.999mA | 0.001mA | ±(1% + 0.002mA) |
Ramp-Up Time | 0~999.9s | 0.1s | ±(1% + 0.1s) |
Ramp-Down Time | 0~999.9s | 0.1s | ±(1% + 0.1s) |
Test Time | 0.5~999.9s | 0.1s | ±(1% + 0.1s) |
Insulation Resistance Test | |||
Range | Resolution | Accuracy | |
Output Voltage | 100~6000V | 1V | ±(1% rdg. + 5 V) |
Insulation Resistance Test | 0.5M~100GΩ | 0.01M/0.1M/1M/10M/100M/1G | 0.60M~9.99M |
±(1% rdg.+0.03M) | |||
10.0MΩ~99.9MΩ | |||
±(3% rdg.+0.3M) | |||
100MΩ~9999MΩ | |||
±(5% rdg.+3M) | |||
10.0GΩ~100GΩ | |||
±(15% rdg.+ 0.8G) | |||
Ramp-Up Time | 0~999.9s | 0.1s | ±(1% + 0.1s) |
Ramp-Down Time | 0~999.9s | 0.1s | ±(1% + 0.1s) |
Test Time | 0.5~999.9s | 0.1s | ±(1% + 0.1s) |
Delay Time | 0~999.9s | 0.1s | ±(1% + 0.1s) |
Control & Interfaces | |||
PLC Digital I/O | Pass, Fail, Testing | ||
RS232 | Standard | ||
WiFi | Standard | ||
* Due to layout limitations, only best‑selling models are shown.Contact us for a complete product selection.